RESET RESERVOIR VOLUME
The reservoir volume setting indicates
the amount of fluid contained in the
reservoir. Once this number is set, the
pump keeps track of how much fluid
has been delivered and adjusts the
reservoir volume setting accordingly.
Note: If you are running a step or taper
therapy, resetting the reservoir volume
also resets the infusion. When you
restart the pump, delivery starts at the
beginning of the infusion duration.
To reset the reservoir volume after
attaching a new cassette:
1.
The pump displays a question
asking if you wish to reset the
volume to the default amount.
If this screen does not appear,
the reservoir volume may already
be reset
2.
If you are running a step or taper
therapy, you also see, “Infusion will
be reset”. Select Yes to reset the
volume and infusion. Select No to
keep the reservoir volume at the
current setting
To reset the reservoir volume without
changing the cassette:
1.
Stop the pump if it is running
2.
In the Tasks menu, press
▲
or
▼
until Reset Reservoir Volume is
highlighted, and then press Select
3.
The pump displays a screen asking
you to confirm that you want to
reset the reservoir volume.
Select Yes