Clinician Guide
p. 30
Prime Tubing After Changing
a Cassette
Note: If you are not changing the
cassette but wish to prime the fluid
path, use the Prime Tubing, No
Cassette Change task.
1.
When a cassette is attached after
the pump is powered on, a “Prime
Tubing?” screen always appears
2.
Select Yes
3.
Unlock the keypad, if required
4.
If you have not already done so,
disconnect the tubing from the
patient, open the clamps and
select Prime
5.
Stop priming at any time by
selecting Stop Priming. Priming
automatically stops after 10 mL
(or 20 mL if a high volume set is
attached) are primed. Continue
priming as needed
Prime Tubing, No Cassette Change
1.
Stop the pump if it is running
2.
In the Tasks menu press
▲
or
▼
until Prime Tubing is highlighted,
and then press Select
3.
Unlock the keypad, if required
4.
If you have not already done so,
disconnect the tubing from the
patient, open the clamps,
and select Prime
5.
Stop priming at any time by
selecting Stop Priming. Priming
automatically stops after 10 mL
(or 20 mL if a high volume set is
attached) are primed. Continue
priming as needed