Low Priority Alarm
A low priority alarm does not stop
the pump if it is running. The alarm
is accompanied by a blue screen,
and it persists for five seconds
unless it is acknowledged by the
press of a key or the condition
that triggered the alarm goes
away before the five seconds
have passed. (Some low
priority alarms persist for
longer than five seconds)
•
Informational Message
An informational priority message
does not stop the pump if it is
running. This message appears in
the status bar, and does not display
a new alarm screen. It persists for
five seconds and may be silent,
requiring no acknowledgment.
Some informational examples
are “Cassette Locked,” and
“Cassette Unlocked”
Alarm Help Screens
Additional information may be
displayed when certain alarms occur.
The help screens describe what you can
do to try to solve the current problem
that is causing the alarm.
1.
When an alarm occurs, select
Silence to quiet the alarm
2.
If help screens are available for
the alarm, “Help” appears above
the right soft key. To view the help
screens, select Help
3.
Follow the applicable instructions
provided on the help screen. To
page through all available help
screens, press Next repeatedly.
Press Acknowledge at any time
to exit Help
4.
“Retry Help” appears when no
additional help steps are available.
To page through the help screens
again, press Retry Help. The Alarm
screen reappears as in Step 2
above. Repeat Steps 2 and 3. When
the alarm clears, the help screens
will no longer be displayed