How to ensure safety
Pay attention to the safety instructions
S
Attention!
Do
not
switch on the HI-SCAN unit, if it has not been operated for a period of
6
months
! The X-ray
generators must first be regenerated by a competent service technician! The non-respect of this rule
may lead to the
destruction of the X-ray generators
!
30
A
Warning!
Store the X-ray unit in
closed rooms only
and do
not
operate it
outdoor
! Only operate the unit if the
operating conditions regarding ambient temperature (
0°C to 40°C
) and humidity of the air (
10% to
90%, not condensing
) are kept during operation! (In units with extended climatic range, please obtain
the operating conditions from the unit's purchase documents.)
Protect
the unit
entirely
against
heavy
dirt
(dust, waste gas etc.) as well as
humidity
(rain, snow etc.)
and
direct sunlight
by appropriate measures! This also applies to units installed in vehicles, special
containers etc. for mobile use!
40
A
Warning!
If other persons are charged with the operation of the HI-SCAN X-ray unit, ensure that they are
com-
petent operators
who know
all the safety instructions
and
legal and official provisions
! Roller and
modular plastic belt conveyors may
only
be loaded and unloaded
by
trained staff
who are
familiar
with the accident prevention regulations
!
50
A
Warning!
Be absolutely sure that the local
mains voltage corresponds
to the value indicated on the X-ray unit
type plate and the fusing of the provided mains outlet meets
at least
the value indicated on the type
plate!
Be also sure that the provided mains outlet is equipped with a
protective earth conductor
! A connec-
tion without PE conductor is
not allowed
!
If you connect the unit to the
on-board supply system
of a vehicle, airplane or ship or to another
spe-
cial mains supply
, pay attention to the prevailing safety instructions!
60
A
Warning!
Make sure that all available
emergency stop keys
(
F
1-12pp.), the
mains plug
and the
main switch*
at the mains input of the X-ray unit or the
mains plug
at the mains outlet are
accessible
during oper-
ation! If required, remove objets that might obstruct the access!
70
A
Warning!
Mechanical and electrical modifications of the X-ray unit may
only
be performed by a
trained service
technician
! The regulations and instructions given in the
Technical Manual
must be adhered to!
80
95587412 01/01/09 © Smiths Heimann
1-5
Содержание HI-SCAN 100100T
Страница 2: ...95584681 01 01 09 Smiths Heimann ...
Страница 4: ...Contents IV 95584681 01 01 09 Smiths Heimann ...
Страница 5: ...95584770 01 03 2003 Smiths Heimann 1 ...
Страница 6: ......
Страница 10: ......
Страница 117: ...95584770 01 03 2003 Smiths Heimann 2 ...
Страница 118: ......
Страница 122: ......
Страница 218: ...95587413 01 01 09 Smiths Heimann 2 96 ...