How to correct operating troubles
Error messages
Cause/Correction
Error code: 0014
Object inside tunnel or signal deviation of a detector
line! Convey object out of tunnel or switch off X-ray
unit and remove object manually!
A
R
Warning!
Switch off the X-ray unit before starting any of
the following work and remove the key! If the
unit does not have a key switch, wait until all
indicators on the keyboard have extinguished
and separate the unit from the mains by
pulling the mains plug
!
1220
When clearing the tunnel after switching on the
X-ray unit, ensure that no baggage or freight
items are placed on the conveyors while the con-
veyors are running.
Check if there is still a baggage item inside the
inspection tunnel and remove it by hand.
If there is no item in the tunnel, you should ask
the local service office or Smiths Heimann to do
the repair, because image errors must be taken
into consideration due to the signal deviation.
Error code: 0018
Operating conditions out of range! Scanning disabled!
The temperature is too high or too low. See to a
lower or higher ambient temperature to be estab-
lished.
Error code: 0027
X-ray system (A/B/...) cannot be activated (operating
conditions are possibly out of range)!
On units equipped with a heating or cooling sys-
tem* for the X-ray generators, the correct operat-
ing conditions will be automatically established
after a short time.
Error code: 0028
X-ray system (A/B/...) cannot be deactivated! Switch
off unit immediately!
Due to a hardware failure the X-radiation does not
switch off anymore.
A
R
Warning!
Switch off the X-ray unit
at once
! If the error
returns after switching on again or if the er-
ror 0029 (
Unexpected activation of X-ray sys-
tem!
) occurs, withdraw the key or the mains
cable and have the unit repaired! Further op-
eration of the X-ray unit is not permitted!
1230
Error code: 0029
Unexpected activation of X-ray system (A/B/...)!
Switch off unit immediately!
Due to a hardware failure X-radiation switches on
by itself.
A
R
Warning!
Therefore switch off the X-ray unit
at once
! If
the error returns after having switched on
again, withdraw the key or the mains cable
and have the unit repaired! Further operation
of the X-ray unit is not permitted!
1240
95587412 01/01/09 © Smiths Heimann
1-97
Содержание HI-SCAN 100100T
Страница 2: ...95584681 01 01 09 Smiths Heimann ...
Страница 4: ...Contents IV 95584681 01 01 09 Smiths Heimann ...
Страница 5: ...95584770 01 03 2003 Smiths Heimann 1 ...
Страница 6: ......
Страница 10: ......
Страница 117: ...95584770 01 03 2003 Smiths Heimann 2 ...
Страница 118: ......
Страница 122: ......
Страница 218: ...95587413 01 01 09 Smiths Heimann 2 96 ...