[F 0] Reference condition/Units selection function/Switch output function
Reference condition
Standard condition or normal condition can be selected.
Standard condition and normal condition are defined as follows:
•Standard condition: Displayed flow rate which is converted to volume at 20°C, 101.3 kPa (absolute pressure).
•Normal condition: Displayed flow rate which is converted to volume at 0°C, 101.3 kPa (absolute pressure).
Units selection function
With the units selection function, the selectable display units are L/min or cfm (ft
3
/min).
This setting is only available for models with the units selection function.
: This function is not displayed for models without unit selection function.
: For the product without the unit selection function, [L] is shown in the sub display on the left.
Switch output type
The switch output function can be selected (NPN or PNP output).
<Operation>
Display [F_0] by pressing
▲
or
▼
button in function selection mode.
Press the S button.
Move on to the reference condition.
Reference condition
Press the
▲
or
▼
button to select the reference condition.
Press the S button to set.
Move on to the units selection function.
Units selection function
Press the
▲
or
▼
button to select the display unit.
Press the S button to set.
Move on to the switch output function.
Switch output function
Press the
▲
or
▼
button to select the switch output function.
Press the S button to set.
Return to function selection mode.
[F 0] Reference condition/Units selection function/Switch output function completed