5 Operation
Operating Instruction SIROSONIC/L/TL, PerioSonic
58
62 78 415 D3275
.
201.04.09
.
02
5.7.6
Using the endo tip no. 5
This ultrasonic tip is intended for flushing the prepared root canals
(invasive procedure). Before treatment, the spray adapter must be
inserted for supplying external irrigation fluids (e.g. NaCl).
Inserting the spray adapter
✔ Cooling water flow is cut off using control ring (A).
1. Detach the ultrasonic handpiece from the supply hose.
2. Insert the spray adapter (B) between the hose coupling and the
handpiece. Note the connecter and hole pattern.
3. Attach the NaCl hose (C) and fasten it to the supply hose with the
hose clamps (D).
6278415.book Seite 58 Montag, 17. Oktober 2016 11:28 11