5 Operation
Operating Instruction SIROSONIC/L/TL, PerioSonic
32
62 78 415 D3275
.
201.04.09
.
02
5.4
Checking the ultrasonic tip for usability
✔ The included checkcard for tips is available.
1. Place the ultrasonic tip against the tip gauge with the edge (A) of the
wrench surface positioned as shown in the corresponding
illustration.
2. Check the ultrasonic tip based on the checkcard. The marking lines
(B) show the maximum permissible wear.
Does the ultrasonic tip deviate from the contour of the illustration on the
checkcard or is the ultrasonic tip shorter than the marking line?
1. Adjust the usage of the ultrasonic tip.
2. Dispose of the ultrasonic tip and associated torque wrench.
The illustrations on the checkcard show the correct tip shapes and
lengths.
Any deviation from the contour shown on the figure represents an
unacceptable distortion of the shape.