Operating Instruction SIROSONIC/L/TL, PerioSonic
7 Conditioning
62 78 415 D3275
.
201.04.09.02
71
1. Brush the ultrasonic handpiece, detached tip and torque wrench
under running water (< 38 °C, < 100 °F, at least drinking water
quality) and good lighting (min. 500 lux) and color rendering index
(min. 80 Ra) until no more dirt can be seen, for at least 10 seconds.
2. Blow spray channels with 2.5 - 3 bar until no more moisture
displaces, but at least 10 seconds.
3. Conduct thermal disinfection or unwrapped steam sterilization [ →
67].
4. Pack the instruments in packaging material suitable for sterilization
and storage. e.g. paper/plastic composite packaging.
5. Conduct sterilization [ → 72].