Operating Instruction SIROSONIC/L/TL, PerioSonic
5 Operation
62 78 415 D3275
.
201.04.09.02
57
Use
1. Place the tip on the inlay without applying any pressure.
2. Switch the handpiece on for a few seconds to greatly reduce the
viscosity of the hardening composite (thixotropy) and insert the inlay
almost without using pressure.
ª
After the handpiece is switched off, the hardening composite
immediately returns to its normal consistency, i.e. excess
material does not flow away and can be easily removed.
3. Repeat the process to check whether any surplus composite
material can be expelled.
ª
Curing takes place according to the composite manufacturer's
specifications.
6278415.book Seite 57 Montag, 17. Oktober 2016 11:28 11