Operating Instruction SIROSONIC/L/TL, PerioSonic
5 Operation
62 78 415 D3275
.
201.04.09.02
33
5.5
Adjusting the power for the ultrasonic tips
CAUTION! Wear and bending of the tips lead to loss of performance
and hazards. Tips may not be reshaped. There is a risk of breakage
during operation and performance may be lost or reduced.
The output power of the scaler tips is approx. 10 W; that of the CEM tip
is approx. 1 W.
IMPORTANT: The user is responsible for setting the appropriate power
level according to the medical indication.
➢ Adjust the power of the ultrasonic tip on the treatment center or
PerioScan. Observe the power settings [ → 22].
Darwin
CAUTION! Endodontic tip no. 5 may be operated only at a power
level of max. 4%! (SIROSONIC/L, PerioSonic)
SINIUS, TENEO
CAUTION! Endodontic tip no. 5 may be operated only after activating
the Endo function (Endo key on the touchscreen of the treatment
center)! Here it is possible to preset 1e to 5e. Please note that these
6278415.book Seite 33 Montag, 17. Oktober 2016 11:28 11