Operating Instruction SIROSONIC/L/TL, PerioSonic
7 Conditioning
62 78 415 D3275
.
201.04.09.02
67
7.2
Automated cleaning and disinfecting ...
7.2.1
... with a DAC UNIVERSAL
Sirona - DAC Universal (Flex lid)
We recommend using Sirona DAC UNIVERSAL (Flex lid) for
automated cleaning and disinfection (inside and outside).
For further details, refer to the operating instructions supplied with the
unit.
Procedure _ Instruments (plural) without pushbutton chuck
✔ The instruments are conditioned using the DAC UNIVERSAL.
1. Check whether the instruments are clean after conditioning under
good lighting (min. 500 lux) and color rendering index (min. 80 Ra).
2. If they are still dirty, repeat the process.
ª
For further conditioning, the instruments should be dry and free
of residue.
3. Pack the instruments in packaging material suitable for sterilization
and storage. e.g. paper/plastic composite packaging.
4. Conduct sterilization [ → 72].
DAC Univ
ersal
DAC Univer
sal
6278415.book Seite 67 Montag, 17. Oktober 2016 11:28 11