4 Preparation
Operating Instruction SIROSONIC/L/TL, PerioSonic
26
62 78 415 D3275
.
201.04.09
.
02
4. Check the ultrasonic tip for usability [ → 32].
5. Set a sufficient amount of cooling water (> 30 ml/min) [ → 35].
6. Use filtered water only (< 50 µm).
7. Check the nozzles for blockages and lime deposits, for example, and
clean the nozzles if necessary.
8. Make sure the ultrasonic tip is fits tightly.
9. Perform an axial load test of the ultrasonic tip.
Tip: Use a cofferdam in patients for endodontic work.