Operating Instruction SIROSONIC/L/TL, PerioSonic
5 Operation
62 78 415 D3275
.
201.04.09.02
35
5.6
Adjusting the cooling spray
General adjustment
> 30 ml/min
➢ Adjust the flow rate of the cooling water using the control ring (A) (>
30 ml/min).
Tip: You can measure the amount of cooling water with a measuring
cup and watch.
Scaler except no. 6L
Exception: The water flow of the CEM tip no. 6L is blocked.
When using CEM tip no. 6L, the ultrasonic handpiece can be operated
only intermittently.
Max. water flow
Water flow rate
Marking logo
... for SIROSONIC und SIROSONIC L:
The water flow is reduced and lastly shut off when the control ring is
turned in a clockwise direction.
The maximum water flow is set when the control ring is rotated left until
perceptible resistance. If the control ring is turned beyond the
resistance, the water flow is reduced until it shuts off.
A
A