7 Conditioning
Operating Instruction SIROSONIC/L/TL, PerioSonic
70
62 78 415 D3275
.
201.04.09
.
02
7.3
Manual cleaning and disinfection
Manual conditioning is possible in exceptional cases
IMPORTANT: Manual conditioning is possible in exceptional cases if
the national/regional regulations are followed. The national/regional
regulations are to be checked before.
Immediately after treatment
NOTICE! Condition immediately, or at the latest, one hour after
treatment.
NOTICE! Never clean in an ultrasound bath!
NOTICE! Never immerse in disinfectant solution!
No strong acidic, alkaline or chlorine solutions
NOTICE! Do not use any strong acidic or alkaline solutions (5 < pH < 9)
or solutions containing chlorine.
Use soft brush
IMPORTANT: Use a soft, clean, and disinfected brush for cleaning.
Procedure _ Scaler
✔ Wear appropriate protective clothing.
✔ All disinfectants must be approved in your country and have proven
bactericidal, fungicidal and virucidal properties. Only use
disinfectants with no protein-fixing properties.