Operating Instruction SIROSONIC/L/TL, PerioSonic
5 Operation
62 78 415 D3275
.
201.04.09.02
37
5.7
Treatment
CAUTION! If the sealing ring of SIROSONIC TL or PerioSonic is
defective, damaged or worn, this can lead to false detection results!
Always check for proper fit and condition of the sealing ring before using
the SIROSONIC TL and PerioSonic.
CAUTION! When the ultrasonic handpiece is not in use, detach the
ultrasonic tip and attach the torque wrench to protect the tip. Make sure
to keep the instrument away from soft tissues such as cheeks, lips,
tongue, etc. during treatment. Contact causes friction that can lead to
burns!
Select the most suitable instrument tip for each treatment.
The optical and acoustic display of calculus removal in connection with
the PerioScan equipment is possible only with instrument tip
SIROPERIO 4 PS (PS 4). For this, the distal end of the tip must always
touch the root surface and may not be too worn.