Operating Instruction SIROSONIC/L/TL, PerioSonic
8 Maintenance
62 78 415 D3275
.
201.04.09.02
77
SIROSONIC TL / PerioSonic
... for SIROSONIC TL or PerioSonic
1. Detach the ultrasonic handpiece from the supply hose.
2. Position the longest partial marking of the control ring (A) opposite
the recess (B). The spray water cartridge can be replaced only in this
position.
3. Remove the spray water cartridge (C) with a tweezers.
4. Spray oil lightly onto the new spray water cartridge.
5. Insert the new spray water cartridge up to the stop.
6278415.book Seite 77 Montag, 17. Oktober 2016 11:28 11