5.5 List of error messages
Service Manual GALILEOS
E1 06 42 GALILEOS
E1 06 43 GALILEOS
E1 06 44 GALILEOS
E1 06 45 GALILEOS
E6 06 51 GALILEOS
E6 06 52 GALILEOS
E6 06 53 GALILEOS
E6 06 54
E6 06 55
E6 06 56
Error code
Description
Actions required
see
E1 06 42
Tolerance exceeded
mA nom.
● Replace the tube assembly.
Error code
Description
Actions required
see
E1 06 43
Tolerance exceeded
VH actual value
● Replace the tube assembly.
Error code
Description
Actions required
see
E1 06 44
Tolerance exceeded kV actual
value
● Replace the tube assembly.
Error code
Description
Actions required
see
E1 06 45
Tolerance exceeded mA actual
value
● Replace the tube assembly.
Error code
Description
Actions required
see
E6 06 51
VHmax
● Replace the tube assembly.
Error code
Description
Actions required
see
E6 06 52
MAmax
● Replace the tube assembly.
Error code
Description
Actions required
see
E6 06 53
KVmax
● Replace the tube assembly.
Error code
Description
Actions required
see
E6 06 54
Basic heating pulses not applied.
● Replace the tube assembly.
Error code
Description
Actions required
see
E6 06 55
Anode voltage too low.
● Replace the tube assembly.
Error code
Description
Actions required
see
E6 06 56
Error during auto-compensation.
● Please contact the Sirona Customer
Service Center (CSC) to find out
whether a bug fix by means of a
software update is possible and
perform such an update if necessary.
● Let the tube assembly cool down for
approx. 30 mins and repeat this
procedure.
If the error occurs repeatedly ...
● Replace the tube assembly.