61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
189
Sirona Dental Systems GmbH
7
Adjusting/calibrating the unit
Service Manual GALILEOS
7.3
Adjustment and calibration via the "Service Functions" menu
båÖäáëÜ
7.3.9
Service
Service
You can create a test image with the
"Service"
tab card.
It is not necessary to execute this menu for the calibration of the unit!
✔
The
"Service functions"
1. Click the
"Service"
tab.
ª
The corresponding tab card is selected.
2. Click on the
"Image acquisition"
button.
ª
SIDEXIS XG makes the unit ready for exposure [ → 162].
ª
Service routine S032.41 is displayed on the control panel.
3. Press the R key to move the unit back to the starting position.
4. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal indicating the end of the exposure
(double beep) sounds (if it has been configured).
ª
The CCD test image is displayed.
Содержание galileos
Страница 1: ...MSKOMNO d ifiblp kÉï ë çÑW båÖäáëÜ pÉêîáÅÉ j åì ä GALILEOS Service Manual ...
Страница 2: ......
Страница 387: ......