Sirona Dental Systems GmbH
6
Troubleshooting
Service Manual GALILEOS
6.1
Error logging memory
båÖäáëÜ
6
Troubleshooting
Troubleshooting
6.1
Error logging memory
Error logging memory
The error logging memory is part of the extended details.
Data which might be expected to occur in the error logging memory is
explained below to aid you in interpreting it.
DANGER
Potentially lethal shock hazard!
It is essential to switch off the unit and wait at least 1 minute before
removing a cover.
Switch OFF the X-ray unit before connecting a measuring instrument.
Perform continuity tests only on units which are switched off.
NOTICE
Risk of damage to unit
Select the correct current/voltage type and adjust the measuring range
to match the expected readings.
Keep to the prescribed cool-off periods if several exposures have to be
taken to check a measured value.
Please observe the usual precautionary measures for handling printed
circuit boards (ESD). Touch a ground point to discharge static electricity
before touching any boards.
CAN bus cable: When unplugging CAN bus cables, it is essential to
unplug the power supply as well.