61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
135
Sirona Dental Systems GmbH
6
Troubleshooting
Service Manual GALILEOS
6.3
Checking the boards
båÖäáëÜ
GALILEOS LEDs
Important LEDs on the boards
(see also wiring diagrams)
*) Starting with unit serial number 3201, new units will be delivered
without board DX41.
Board
LEDs
Normal operation
Malfunction
Bootloader
DX1
V100
lit
not lit
V101
lit
not lit
V108
lit
not lit
V110
lit
not lit
V200
flashing at 1 Hz
lit or
not lit
V610
lit
not lit
flashing at 2 Hz
DX6
V1
flashing at 1 Hz
not lit
flashing at 2 Hz
V203
lit
not lit
DX7
V100
lit
not lit
V101
lit
not lit
V102
lit
not lit
DX71
V101
lit
not lit
V103
lit
not lit
V107
flashing at 1 Hz
not lit
flashing at 2 Hz
DX32
V132
lit
not lit
V133
lit
not lit
DX41*
V103
flashing at 1 Hz
not lit
flashing at 2 Hz
V202
lit
not lit
V204
lit
not lit
DX42
V101
lit
not lit
V103
lit
not lit
DX89
V201
lit
not lit
V202
lit
not lit
V203
lit
not lit
V204
flashing at 1 Hz
not lit
flashing at 2 Hz
V205
lit
not lit
V207
lit
not lit
Содержание galileos
Страница 1: ...MSKOMNO d ifiblp kÉï ë çÑW båÖäáëÜ pÉêîáÅÉ j åì ä GALILEOS Service Manual ...