61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
71
Sirona Dental Systems GmbH
4
General operating procedures
Service Manual GALILEOS
4.3
Configuring the unit
båÖäáëÜ
4.3
Configuring the unit
Configuring the unit
4.3.1
Configuring the unit via the SIDEXIS Manager
Configuring the unit via the SIDEXIS Manager
SIDEXIS configuration: GALILEOS
The X-ray component must be set up and enabled using the SIDEXIS
Manager. For more information, refer to SIDEXIS XG Installation
Instructions.
4.3.2
Configuring Facescan
Configuring Facescan
Facescan is generally configured via the integrated web dialog on the
FaceScan unit [ → 75].
There are however two basic options for configuring the Facescan unit:
● Type 1: Configuration using the Facescan USB stick [ → 71]
● Type 2: Configuration over a network cable (peer-to-peer) [ → 75]
4.3.2.1
Type 1: Configuration using the Facescan USB stick
Type 1: Configuration using the Facescan USB stick
Connecting the USB stick
1. Pull gray cable L78.4 from slot X2 of the FACESCAN modular board.
2. Unscrew the protective plate (T) from the Facescan unit.
ª
The FACESCAN modular board is visible.
3. Plug cable L78.4 again into slot X2 on the FACESCAN modular
board.
Содержание galileos