9 Repair
Sirona Dental Systems GmbH
9.6 X-ray tube unit
Service Manual GALILEOS
With the "Type 1" diaphragm unit:
1. Remove the retaining ring (L).
2. Remove the copper filter (M) and the aluminum filter (D).
3. Unscrew the two sleeves (C) with washers.
4. Loosen the two lower screws (A) (approx. 2 to 3 turns).
With the "Type 2" diaphragm unit:
1. Remove the retaining ring (L).
2. Remove the copper filter M and the aluminum filter (D).
With the "Type 3" diaphragm unit:
➢ No preparations are required.
Installing the new tube assembly
1. Hang the new tube assembly on the two front screws of the rotating
element and tighten them securely.
2. Insert the two rear screws and tighten them firmly.
3. Plug cables L3, L5, L6 and L15 as well as the ground cable back onto
board DX6 and reattach the cables to the rubber grommets.
4. Reattach the cover plate.