Repair
Service Manual GALILEOS
9.13
Cable
båÖäáëÜ
3. Screw the new energy chain down in the position (marking) of the old
energy chain.
4. Screw down the mains filter plate.
5. Screw the power cable (A) to the terminal K1 and the strain relief.
6. Lead the external PE cable (B) over the mains filter plate through the
ferrite core twice (C) (a hose).
7. Screw the external PE cable (B) down on the ground bolts (D).
IMPORTANT
Possible assembly errors
The one end piece has been removed to make the new energy chain
easier to lay .
The missing end piece is enclosed.
➢ Attach this end piece to the energy chain only when the energy
chain has been laid in the stand.