61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
177
Sirona Dental Systems GmbH
7
Adjusting/calibrating the unit
Service Manual GALILEOS
7.3
Adjustment and calibration via the "Service Functions" menu
båÖäáëÜ
ª
The lower edge of the lower lead diaphragm must be within
tolerance, i.e. lie inside of the auxiliary lies (A).
If the edge is out of tolerance (B), the diaphragm must be
adjusted mechanically [ → 197].
5. If the exposure is OK (A), confirm this by clicking the check box
located to the right of the
"Lower jaw"
button.
ª
The box will appear checked.
ª
Diaphragm adjustment is now complete.
6. Continue the calibration procedure with the radiation field
A
Adjustment OK
B
Adjustment not OK