61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
247
Sirona Dental Systems GmbH
8
Service
Service Manual GALILEOS
8.9
S011: Dosimetry (without ring movement)
båÖäáëÜ
8.9
S011: Dosimetry (without ring movement)
S011: Dosimetry (without ring movement)
S011: GALILEOS
* SR=service routine, ** SHZ=security access
8.9.1
S011: Test step 12
S011: Test step 12
Dosimetry with pulsed radiation
1. Call service routine S011.12 [ → 213].
ª
Selection field 1 displays "85kV/42mAs".
WARNING
Unit is radiating X-rays
Excess exposure to X-rays is detrimental to health.
➢ Use the prescribed accessories for radiation protection.
➢ Do not stay in the X-ray room during exposure. Move as far away
from the unit as the coiled cable for the release button allows you to.
SR*
SHZ**
Function
S011
Dosimetry (without ring movement)
S011.9
Yes
4s Continuous radiation with 85kV/7mA (for current measurement)
S011.12
Yes
Dosimetry with pulsed radiation
Содержание galileos
Страница 1: ...MSKOMNO d ifiblp kÉï ë çÑW båÖäáëÜ pÉêîáÅÉ j åì ä GALILEOS Service Manual ...
Страница 2: ......
Страница 387: ......