Instruction Manual ColorPlus 3
Servicing
66
14501E/2
9.7
Changing the battery in the SICON (M)
DANGER!
Life-threatening voltage inside the instrument.
Connecting electrical lines can be extremely dangerous. Instrument parts may also be dam-
aged. Local regulations for electrical installations must be observed at all times.
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Interrupt the service voltage to the SICON (M).
2.
Open the SICON (M) according to Section 5.2.
3.
Remove the battery (circle).
4.
Insert the new battery.
5.
Close the SICON (M).
6.
Re-establish the service voltage.
7.
Set the date and time.