Servicing
Instruction Manual ColorPlus 3
14501E/2
61
WORKSTEP
ADDITIONAL INFO / IMAGES
5.
Clean the four measuring cell windows (ar-
rows) with ethanol.
Do not use abrasive cleaning agents.
6.
Remove the remaining part of the measuring
cell.
7.
Grease the gaskets on the sample distributor
and outlet pipe (position X).
Check the gasket on the sample dis-
tributor (8 x 1) and outlet pipe (6 x 1) and
replace if necessary.
Insert the sample distributor (B) and outlet
pipe (A) back in the bores.
Ensure that the pins are aligned to the
slots (see C) and that the outer diameters of
the sample distributor and outlet pipe are
different.
8.
Screw the cap and screw cap onto the measur-
ing cell.