Instruction Manual ColorPlus 3
Servicing
62
14501E/2
9.3
Replacing the desiccant
The following procedure describes how to replace the desiccant in the ColorPlus 3:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Unscrew the cover on the ColorPlus 3.
2.
Replace the old desiccant bag with a new one.
Check the gasket on the cover (88 x 2).
Replace this, if necessary.
3.
Screw the cover back onto the ColorPlus 3
immediately.
The humidity sensor is located in the
receiver. Therefore, the humidity measure-
ment only reacts slowly to new desiccant
(2% to 3% per hour).