Instruction Manual ColorPlus 3
Servicing
60
14501E/2
9.2
Cleaning the measuring cell
The following procedure describes how to clean the measuring cell and measuring cell win-
dows on the ColorPlus 3:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Stop the sample flow and let the measuring
cell run dry.
2.
Interrupt the service voltage to the photome-
ter.
Eye injuries caused by UV radiation.
The instrument uses a xenon flash lamp. The
light pulses are intense and the wavelength
spectrum extends from UV to IR.
Always interrupt the service voltage be-
fore opening the instrument.
Never look directly into the light beam.
Do not use any reflective objects when
cleaning the measuring cell.
3.
Loosen the screw cap and remove it from the
ColorPlus 3 together with the cap.
4.
Pull out the sample distributor (A) and outlet
pipe (B) and clean with a bottle brush.