Instruction Manual ColorPlus 3
Commissioning
36
14501E/2
6
Commissioning
The initial start-up of the web user interface via the Ethernet interface is described in the
Reference Manual. If malfunctions occur, consult the Section 10.
Proceed with the initial start-up in accordance with the following table:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Ensure that the photometer and control unit
are correctly mounted and connected.
2.
Establish the service voltage to the SICON.
The welcome screen appears on the display.
The factory setting language is English.
2.2: The instrument carries out an internal
functional check.
2.3: The instrument is ready for measurement.
3.
Set the language.
Section 8.1
4.
Set the current outputs.
Section 8.2
5.
Set the limits.
Section 8.3
6.
Set the outputs.
Section 8.4
7.
Activate the optional flow meter.
Section 8.5
8.
Set the date and time.
Section 8.6
9.
Enter the access code.
Section 8.7
10. Carry out recalibration.
Section 9.4
11. Back up the configured data.
Section 8.8