Instruction Manual ColorPlus 3
Mounting
20
14501E/2
4
Mounting
4.1
Mounting the ColorPlus 3
Note the following points when mounting the instrument:
The ColorPlus 3 must be installed horizontally. The sample outlet (Y) must be positioned
at the top so that the measuring cell can be well ventilated.
A regulator valve must be installed on the sample outlet (Y) so that counter pressure can
be generated (operation of the measuring cell under pressure so that disruptive air bub-
bles are eliminated).
Mount the ColorPlus 3 as follows:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Install the ColorPlus 3 horizontally onto a wall
using two M6 screws on the fastening plate.
2.
Connect the hoses (D = 8 mm) to the sample
inlet and outlet.
X: Sample inlet
Y: Sample outlet
The G ¼" threads can be accessed by
removing the push-in connections (X, Y).