Instruction Manual ColorPlus 3
Instrument overview
16
14501E/2
Measuring cell data
Values
Material
PVC housing, stainless steel 1.4435
Window
Quartz glass
Plastic push-in connec-
tions
D = 8 mm / G ¼" thread
Max. medium tempe-
rature
0 .. 50 °C
Max. medium pressure 600 kPa (6 bar)
Flow rate
0.5 .. 1 l/min
SICON control unit
Values
Service voltage and
power consumption
24 VDC ± 10 %
5 W (only SICON)
Display
¼ VGA with touchscreen
Resolution: 320 x 240 pixels with 3.5" diagonal
Outputs/inputs
Outputs:
4 x 0/4 .. 20 mA, galvanically isolated up to max. 50 V relative
to ground and max. 500 Ω burden.
7 x digital outputs up to max. 30 VDC, freely configurable, 1
output as de-energized closed relay.
Inputs:
5 x digital inputs up to max. 30 VDC, freely configurable.
Interfaces
Ethernet, SD card (for logging, SW update, diagnostics), Modbus
TCP, optional: Modbus RTU, Profibus DP, HART, Profinet IO
Protection class
IP66
Weight
Approx. 0.6 kg
Dimensions
160 x 157 x 60 mm
Housing material
ABS
24 VDC mains device
Values
Service voltage
100 .. 240 VAC, 47 .. 63 Hz
Power consumption
Max. 25 W (the power consumption of the connected sensors
must not exceed 21 W)
Maximum operating
altitude
2,000 m above sea level
Protection class
IP66
Weight
0.66 kg
Dimensions
Approx. 130 x 155 x 55 mm (W x H x D)
Housing material
PC