Electrical installation
Instruction Manual ColorPlus 3
14501E/2
27
5.5.1
Cable cross-sections over longer distances
For connections longer than the standard cable length, an optional connection box must
be connected between the photometer and control unit.
The maximum distance between the control unit and photometer depends on the cable
cross-section used and the available power supply (see table below).
Shielded cables should be used.
Figure 5: Arrangement of control unit over longer distances
Photometer
Connection box
Control unit
The maximum distance (X) between the control unit and connection box depends on the
voltage in the SICON and the cable cross-section used:
Cable cross-section
Max. distance
Remarks
[mm2]
[m]
0.14
50
0.25
90
0.34
120
Standard version
0.50
180
0.75
270
1.00
350
1.50
500