6-16HS
7 Malfunctions and error correction
Version 06/2012, Rev. 1.0 • sb-kue
51
Translation of the original operating manual
Pos: 162 /10 0 Sigm a/10 0 BA Z entrif uge n (St anda rd mod ule)/ 070 Störu nge n u nd Fehle rsuc he/0 70- 002 0 Tab elle d er Fehl erm eldu nge n @ 7 \mo d_1 306 735 705 818_ 68. docx @ 43 792 @ 22 @ 1
7.2
Table of error codes
Error no. Kind of error
Measures
Note
1-9
System error
•
Allow to slow down
• Power off/on
All these errors stop the
centrifuge or cause it to
decelerate brakeless
10-19
Speedometer error
•
Allow to slow down
• Power off/on
20-29
Motor error
• Power off
• Ensure ventilation
30-39
EEPROM error
•
Allow to slow down
• Power off/on
With error 34, 35, and 36, the
centrifuge will stop;
with error 37 and 38 only an
error message will be given
40-45
Temperature error
(only for refrigerated
centrifuges)
•
Allow to slow down
• Power off
•
Allow to cool down
•
Provide better ventilation (only air cooled
centrifuges)
•
Provide sufficient water throughput (only
water cooled centrifuges)
46-49
Imbalance error
•
Allow to slow down
• Power off
•
Eliminate the imbalance
50-59
Lid error
• Press lid key
• Close lid
•
Remove foreign matter from the opening of
the lid lock device
With error 50 and 51, the
centrifuge will stop
60-69
Process error
•
Allow to slow down
• Power off/on
With error 60 message
"power failure during run",
with error 61, the message
"stop after power on"
70-79
Communication error
•
Allow to slow down
• Power off/on
80-89
Parameter error
• Power off
•
Allow to cool down
•
Provide for better ventilation
With error 83, error message
only
90-99
Other errors
• Check connections
•
Provide sufficient water throughput (only
water cooled centrifuges)
NOTE
If it is impossible to eliminate the errors, contact the service!
Pos: 163 /01 0 Univ ersal mod ule/ Sei tenwe chsel @ 0\ mod _12 021 1624 431 2_0 .docx @ 1 05 @ 1 @ 1