6-16HS
6 Using the centrifuge
36
Version 06/2012, Rev. 1.0 • sb-kue
Translation of the original operating manual
Progr.: program selection list
This field in the standard menu shows the program that is currently loaded.
When the field is selected, the program list is displayed (for information on
how to work with the programs, chapter 6.3.3 - "
Program mode
").
The program „RAPID_TEMP“ cannot be deleted (see below).
Abb. 14: Program selection list
Programm "ASTM oil"
This program is stored in the system as a default setting of the centrifuge. It
is specifically adapted to the petroleum rotor 11165 with buckets of the
13113 type.
• Select the option "Progr" in the "Standard" menu and confirm the input.
The program selection list will be displayed.
• Select the program "ASTM oil" from the list and confirm the input. The
following values will be loaded:
Language:
English
Rotor/bucket:
11165/13113
RCF:
650
Temperature:
+60°C
Runtime:
12:00 min
Acceleration:
6
Deceleration:
6
In addition, the standstill heating will be activated and the automatic lid
opening function will be disabled.
Program "RAPID_TEMP"
Preheating at a standstill may distort the measurement results and
subsequently cause increased wear of the mechanical components. This is
why the centrifuge has a special program that preheats the centrifuge
rapidly under defined conditions:
• Select the option “Progr” in the “Standard” menu and confirm the
selection. The program list will be displayed.
• Select the program "RAPID TEMP" on the program list and confirm the
selection. The display shows of the maximum rotor speed and the