6-16HS
6 Using the centrifuge
Version 06/2012, Rev. 1.0 • sb-kue
35
Translation of the original operating manual
Rotor: rotor selection list/ automatic rotor identification
This field shows the rotor that is currently being used.
• Select the field "Rotor" and confirm the selection. A list with all of the
possible rotors without buckets is displayed.
• Select the desired rotor. All of the possible rotor/bucket combinations
and additional information for each combination will be displayed.
• Select a rotor/bucket combination and confirm the selection. Press the
function knob in order to accept the data.
Fig. 13: Rotor selection list
Identifying and adapting incorrectly set rotors
The centrifuge automatically identifies the rotor that is being currently used.
• If the system identifies a different rotor than the one that is set, if there
are no different buckets for this rotor, the rotor input will be adapted
automatically. The system will not display a message.
• If the system identifies a different rotor than the one that is set, and if
there are different rotor/bucket combinations for this rotor, the system
will automatically select the rotor with the lowest speed. The system will
display a corresponding message so that the rotor can be selected
manually.
• If the system cannot identify the rotor, a message will be displayed. The
rotor cannot be used in the centrifuge.
This prevents the maximum permissible speed from being exceeded.