Operation
Positioning the Patient
0.0
syngo CT 2007E
Emotion Operator Manual
02/2008
0.0
D.3-5
0.0
Patient weight
D.3
Patients weighing up to 140 kg (300 lbs) can be examined with-
out any restriction.
D.3
Patients weighing up to 200 kg (440 lbs) can be positioned on
the table and examined. The stability of the patient table is not
at risk. Yet, the correct functioning of the table (lift, positioning
precision) cannot be guaranteed 100%.
D.3
✧
Be especially careful when positioning heavy patients on
the table.
✧
Before you start the examination convince yourself that the
patient is not endangered either by the movement of the
table or by the tilt of the gantry.
Problematic patients
D.3
Special caution is required with obese, unconscious, apathetic,
unresponsive or pediatric patients.
D.3
Содержание SOMATOM Emotion syngo CT 2007E
Страница 4: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 ii ...
Страница 8: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 vi ...
Страница 44: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 General Safety Information Safety A 1 22 ...
Страница 59: ...Safety Information about Personal Safety 0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 A 2 150 0 A 2 ...
Страница 172: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Software Related Safety Information Safety A 4 78 ...
Страница 198: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Scan Unit Gantry System Description B 2 16 ...
Страница 240: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Console System Description B 5 24 ...
Страница 246: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Storage Media and Drives System Description B 6 6 ...
Страница 264: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 System Handling C 2 ...
Страница 268: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Operation of the Gantry System Handling C 1 4 ...
Страница 302: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Use of Accessories System Handling C 4 14 ...
Страница 334: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Switching the System On and Off Operation D 1 30 ...
Страница 340: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Function Tests Operation D 2 6 ...
Страница 426: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Maintenance E 2 ...
Страница 442: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Quality Assurance F 2 ...
Страница 450: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 General Information Quality Assurance F 1 8 ...
Страница 460: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Daily Quality Measurements Quality Assurance F 2 10 ...
Страница 520: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Low Contrast Test Optional Quality Assurance F 5 14 ...
Страница 526: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Camera Test Quality Assurance F 6 6 ...
Страница 536: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Technical Specifications G 2 ...
Страница 538: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Standards and Statutory Regulations H 2 ...
Страница 540: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Release Information J 2 ...
Страница 542: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Safety Instructions K 2 ...
Страница 544: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Forms L 2 ...
Страница 546: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 DIP Report M 2 ...
Страница 548: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Location of labels N 2 ...
Страница 566: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Glossary O 18 ...
Страница 574: ...0 0 syngo CT 2007E Emotion Operator Manual 02 2008 0 0 Index P 8 ...