7.4.5
Calibration schedule (4.5.)
Note
•
Four sets of parameters allow you to monitor the Device/Sensor Lifetimes and set up
Maintenance/Service schedules, based on operating hours instead of a calendar-based
schedule, see also Remaining device lifetime (4.2.) (Page 121), Remaining sensor
lifetime (4.3.) (Page 124), and Service schedule (4.4.) (Page 126).
•
Performing a reset to Factory Defaults will reset all the Maintenance Schedule
parameters to their factory defaults.
•
The device operates in years. To view Calibration Interval parameters in hours or days
(via SIMATIC PDM only) see Calibration interval (4.5.1.) (Page 129).
Factory setting:
years
Setting range
1)
:
Hours, days, years
Purpose:
The device tracks calibration intervals based on operating hours and moni-
tors the predicted lifetime to the next calibration. You can modify the Total
Calibration Interval, set schedules for maintenance reminders, and
acknowledge them.
Description:
To access these parameters via SIMATIC PDM:
•
Open the menu Device – Maintenance and select the Calibration
Schedule tab.
•
After modifying values/units as required, click on Write to accept the
change, and Read to view the effect of the change.
•
Click on Calibration Performed to restart the timer and clear any fault
messages.
1)
Selectable only via SIMATIC PDM.
7.4.5.1
Calibration interval (4.5.1.)
Note
The device always operates in years. Changing the units affects only the parameter view of
the Calibration Interval parameters in SIMATIC PDM.
Factory setting:
1.0 year
Setting range:
0 to 20 years
Purpose:
User-configurable recommended time between product inspections.
Description:
Units: hours, days, years, are selectable only via SIMATIC PDM.
Содержание SITRANS LR200 7ML5422
Страница 185: ...SITRANS LR200 with mA HART Operating Instructions 05 2019 A5E32337676 AE 185 B 5 4 PFTE rod antenna threaded connection ...
Страница 193: ...SITRANS LR200 with mA HART Operating Instructions 05 2019 A5E32337676 AE 193 ...
Страница 194: ...SITRANS LR200 with mA HART 194 Operating Instructions 05 2019 A5E32337676 AE ...
Страница 212: ...Index SITRANS LR200 with mA HART 212 Operating Instructions 05 2019 A5E32337676 AE ...