![Siemens SITRANS LR200 7ML5422 Скачать руководство пользователя страница 95](http://html.mh-extra.com/html/siemens/sitrans-lr200-7ml5422/sitrans-lr200-7ml5422_operating-instructions-manual_384663095.webp)
SITRANS LR200 with mA/HART
Operating Instructions, 05/2019, A5E32337676-AE
95
Activation of Reminders
Reminder 1 before Calibration (Required)
Reminder 2 before Calibration (Demanded)
Electronic Temperature
Electronic Temperature
Lowest Value
Highest Value
Wear
Powered days
Poweron Resets
Communication
Security
Remote access
Write protect
Access control
Local access
Local write protected
PIN to unlock
6.3.3
Operating via FDT (Field Device Tool)
FDT is a standard used in several software packages designed to commission and maintain
field devices. Two commercially available FDTs are PACTware and Fieldcare.
Functionally FDT is very similar to PDM. See Operating via SIMATIC PDM (Page 54) for
more detail.
●
To configure a field device via FDT you need the DTM (Device Type Manager) for the
device.
●
To configure a field device via SIMATIC PDM, you need the EDD (Electronic Data
Description) for the device.
6.3.3.1
Device Type Manager (DTM)
A DTM is a type of software that ’plugs into’ FDT. It contains the same information as an
EDD but an EDD is independent of the operating system.
Содержание SITRANS LR200 7ML5422
Страница 185: ...SITRANS LR200 with mA HART Operating Instructions 05 2019 A5E32337676 AE 185 B 5 4 PFTE rod antenna threaded connection ...
Страница 193: ...SITRANS LR200 with mA HART Operating Instructions 05 2019 A5E32337676 AE 193 ...
Страница 194: ...SITRANS LR200 with mA HART 194 Operating Instructions 05 2019 A5E32337676 AE ...
Страница 212: ...Index SITRANS LR200 with mA HART 212 Operating Instructions 05 2019 A5E32337676 AE ...