![Siemens SITRANS LR200 7ML5422 Скачать руководство пользователя страница 169](http://html.mh-extra.com/html/siemens/sitrans-lr200-7ml5422/sitrans-lr200-7ml5422_operating-instructions-manual_384663169.webp)
SITRANS LR200 with mA/HART
Operating Instructions, 05/2019, A5E32337676-AE
169
A.15.5
Sanitary rod antenna
2”, 3”, and 4” sanitary fitting ferrule with integral gasket, with rod antenna
①
Sanitary fitting ferrule
A.15.6
Waveguide antenna
•
You can connect a maximum of two waveguides together.
•
This option is recommended only for clean liquids, and only on vessels without an
agitator, with no turbulence.
•
Horizontal stress on this antenna must be avoided, otherwise mechanical support may be
required.
•
Process temperature and pressure capabilities are dependent upon information on the
process device tag. The reference drawing listed on the tag can be downloaded from our
website at: www.siemens.com/LR200.
①
Vent hole
②
Minimum 80 mm (3")
•
Recommended for products with a dK lower than 3.
•
For the related propagation factor, see Propagation factor (2.8.3.)
(Page 111).
Содержание SITRANS LR200 7ML5422
Страница 185: ...SITRANS LR200 with mA HART Operating Instructions 05 2019 A5E32337676 AE 185 B 5 4 PFTE rod antenna threaded connection ...
Страница 193: ...SITRANS LR200 with mA HART Operating Instructions 05 2019 A5E32337676 AE 193 ...
Страница 194: ...SITRANS LR200 with mA HART 194 Operating Instructions 05 2019 A5E32337676 AE ...
Страница 212: ...Index SITRANS LR200 with mA HART 212 Operating Instructions 05 2019 A5E32337676 AE ...