![Siemens SITRANS LR200 7ML5422 Скачать руководство пользователя страница 123](http://html.mh-extra.com/html/siemens/sitrans-lr200-7ml5422/sitrans-lr200-7ml5422_operating-instructions-manual_384663123.webp)
SITRANS LR200 with mA/HART
Operating Instructions, 05/2019, A5E32337676-AE
123
7.4.2.5
Reminder 1 (required) (4.2.5.)
Factory setting:
0.164 years
Setting range:
0 to 20 years
Purpose:
If Remaining lifetime (4.2.3.) (Page 122) is equal to or less than this value,
the device generates a Maintenance Required reminder.
Description:
1.
Modify values as required.
2.
Set Reminder activation (4.2.4.) (Page 122) to the desired option.
7.4.2.6
Reminder 2 (demanded) (4.2.6.)
Factory setting:
0.019 years
Setting range:
0 to 20 years
Purpose:
If Remaining lifetime (4.2.3.) (Page 122) is equal to or less than this value,
the device generates a Maintenance Demanded reminder.
Description:
1.
Modify values as required.
2.
Set Reminder activation (4.2.4.) (Page 122) to the desired option.
7.4.2.7
Maintenance status (4.2.7.)
Indicates which level of maintenance reminder is active.
In SIMATIC PDM, open the menu Diagnostics > Device Diagnostics, click on the
Maintenance tab, and check the Calibration Schedule Status window.
7.4.2.8
Acknowledge status (4.2.8.)
Indicates which level of maintenance reminder has been acknowledged.
In SIMATIC PDM, open the menu Diagnostics > Device Diagnostics, click on the
Maintenance tab, and check the Calibration Schedule Status window.
7.4.2.9
Acknowledge (4.2.9.)
Purpose:
Acknowledges the current maintenance reminder.
Description:
To acknowledge a reminder via SIMATIC PDM:
•
Open the menu Diagnostics > Device Diagnostics and click on the tab
Maintenance.
•
In the Calibration Schedule section, click on Acknowledge Warnings.
To acknowledge a reminder via the local control buttons or hand-held program-
mer:
•
Press RIGHT arrow twice to open parameter view and activate Edit Mode.
•
Press RIGHT arrow to acknowledge the reminder.
Содержание SITRANS LR200 7ML5422
Страница 185: ...SITRANS LR200 with mA HART Operating Instructions 05 2019 A5E32337676 AE 185 B 5 4 PFTE rod antenna threaded connection ...
Страница 193: ...SITRANS LR200 with mA HART Operating Instructions 05 2019 A5E32337676 AE 193 ...
Страница 194: ...SITRANS LR200 with mA HART 194 Operating Instructions 05 2019 A5E32337676 AE ...
Страница 212: ...Index SITRANS LR200 with mA HART 212 Operating Instructions 05 2019 A5E32337676 AE ...