SITRANS LR200 with mA/HART
122
Operating Instructions, 05/2019, A5E32337676-AE
7.4.2.1
Lifetime expected (4.2.1.)
Note
The device always operates in years. Changing the units affects only the parameter view of
the Remaining Device Lifetime parameters in SIMATIC PDM.
Factory setting:
10.00 years
Setting range:
0 to 20 years
Purpose:
Allows you to override the factory default.
Description:
Units: hours, days, years, are selectable only via SIMATIC PDM.
7.4.2.2
Time in operation (4.2.2.)
Read only. The amount of time the sensor has been operating.
7.4.2.3
Remaining lifetime (4.2.3.)
Read only. Lifetime expected (4.2.1.) (Page 122) less Time in operation (4.2.2.) (Page 122).
7.4.2.4
Reminder activation (4.2.4.)
Factory setting:
OFF
Setting range:
Reminder 1 (Maintenance Required), Reminder 2 (Maintenance Demand-
ed), Reminders 1 and 2, OFF
Purpose:
Allows you to enable a maintenance reminder.
Description:
To modify this parameter via SIMATIC PDM:
•
Go to menu Device – Maintenance.
Содержание SITRANS LR200 7ML5422
Страница 185: ...SITRANS LR200 with mA HART Operating Instructions 05 2019 A5E32337676 AE 185 B 5 4 PFTE rod antenna threaded connection ...
Страница 193: ...SITRANS LR200 with mA HART Operating Instructions 05 2019 A5E32337676 AE 193 ...
Страница 194: ...SITRANS LR200 with mA HART 194 Operating Instructions 05 2019 A5E32337676 AE ...
Страница 212: ...Index SITRANS LR200 with mA HART 212 Operating Instructions 05 2019 A5E32337676 AE ...