
Replacing boards / replacing components
6 - 25
Siemens AG
SPR2-130.061.02
Page 25 of 28
SIREMOBIL Compact
Medical Solutions
Rev. 11
09.04
CS PS 24
System Manual
•
Configure the following parameters in User setup for MEMOSKOP:
•
Perform the IQ quick test.
Operating mode Parameters
Change the programmed parameters only upon cus-
tomer request.
VCR output
Configure to monitor A.
Hardcopy Release
For SIREMOBIL Compact with 1 monitor:
configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
Configure to monitor B.
Text monitor
For SIREMOBIL Compact with 1 monitor:
configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
configure to both monitors.
Split screen
Configure horizontally.
Configure vertically only upon customer request.
Display of time
Configure Yes.
Set Date Time
Set the current date and time.
Hospital Name
Program the hospital.
Storage of Dose on Disk
If area dose product meter is present, (Diamentor) con-
figure Yes. Otherwise configure No.
Dose on hardcopy output
If area dose product meter is present, (Diamentor) con-
figure Yes. Otherwise configure No.