SIREMOBIL Compact
SPR2-130.061.02
Page 28 of 30
Siemens AG
System Manual
Rev. 11
09.04
CS PS 24
Medical Solutions
5 - 28
Adjustments / Programs
Create External disk (1)
MEMOSKOP CE and MEMOSKOP CE-100:
Program the value to "No".
All other types of MEMOSKOP:
If a MOD drive is implemented and/or the option DICOM
Bridge is installed, program the value to "Yes".
Enable Bridge (1)
MEMOSKOP CE and MEMOSKOP CE-100:
Program the value to "No".
All other types of MEMOSKOP:
If a MOD drive is implemented and/or the option DICOM
Bridge is installed, program the value to "Yes".
Default programs change-
able
Default: Yes
The organ programs can be protected against changes by
setting "No".
Default LUT
After switch-on, the programmed brightness/contrast
curve (LUT) is selected automatically.
Default: LUT 2
Default LUT for Subtraction
After switch-on, the programmed brightness/contrast
curve (LUT) is selected automatically for the subtraction
mode, when equipped with a MEMOSKOP C-SUB.
Default: LUT 1
Default Edge
After switch-on, the programmed edge enhancement is
selected automatically. Default: Edge 1
Software backward compa-
tible
Not implemented "Must be set to No".
Unit serial number
Enter the serial number of the SIREMOBIL Compact if it is
equipped with a Dicom bridge. The serial number is
required for identifying the images transferred. Please
note that you have to omit the leading 0 when entering the
serial number.
Disable doserate selection
in user Setup
Set to "Yes" for SIREMOBIL Compact without an area
dose product measurement device.
Set to "No" for SIREMOBIL Compact with an area dose
product measurement device.
Subtraction K-factor,
phase A
Noise suppression, phase A, is programmed in the SUB-
mode.
Phase A: Mask acquisition.
Subtraction K-factor,
phase B
Noise suppression, phase B, is programmed in the SUB
mode.
Phase B: Acquisition of fill image.
Subtraction Time of phase A
The time of phase A in the SUB-mode is programmed.
Phase A: Mask acquisition.
Roadmap K-factor, phase A
Noise suppression, phase A, is programmed in the road-
map mode.
Phase A: Mask acquisition.
(1) in menu Diagnostics
Function:
Comments: