SIREMOBIL Compact
SPR2-130.061.02
Page 22 of 30
Siemens AG
System Manual
Rev. 11
09.04
CS PS 24
Medical Solutions
5 - 22
Adjustments / Programs
•
Switch the SIREMOBIL Compact OFF and ON again.
•
Check the adjustments again. In the 0° position, the object image must correspond to the
one in Fig. 3.
Calibrating the area dose product measurement device
5
Calculating the CONSTANT parameter (Qt * Kg)
5
•
Read and record the Kg value shown on the electronics unit of the area dose product
measurement device.
•
Read the Qt value on the dose chamber of the area dose product measurement device.
•
Multiply these values and record them. The resulting value is the CONSTANT (QT*Kg)
value to be programmed.
Programming
5
•
Connect the service PC to the serial interface on board D1.
•
Start the service program.
•
Select the Adjustment, Parameter menu.
•
In the Combo box "Parameter Groups", select "Diamentor".
•
Click on "Get from Unit".
•
The Parameter "1. Diamentor Existing" must be on YES .
•
Select parameter "2. Constant (Qt * Kg)".
•
In "Value Actual:" field, enter the parameter you have calculated.
•
Click on "Set Value".
•
Click on "Put to Unit" .
•
Click on "Save to File".
Checking the programmed values
5
Prerequisites
The Qt * Kg value was checked and is programmed. Refer to Programming.
•
Connect the service PC to the serial interface on board D1.
This adjustment is used to calibrate the area dose product mea-
surement device of the SIREMOBIL Compact display.
You must exit the service program with "Logoff" and "Quit" to
store the modified parameters to disk as a new file.
Checking the programmed values will activate the self-test in the
electronic system of the area dose product measurement device.
This serves to recalibrate the electronic system of the area dose
measurement device.
NOTE
NOTE
NOTE