Adjustments / Programs
5 - 27
Siemens AG
SPR2-130.061.02
Page 27 of 30
SIREMOBIL Compact
Medical Solutions
Rev. 11
09.04
CS PS 24
System Manual
Move text positions
Allows you to configure the position of superimposed text.
Date / Time format
European: TT MM JJ / HH MM
US:
MM DD YY / HH MM
System configuration
Allows you to read out the current Memoskop configura-
tion.
Keyboard
Allows you to program the corresponding keyboard layout
Universal / Roman / English.
Graphics level
Level (brightness) of superimposed text.
(Default: 128)
Display of real time
clock
Yes: Real time clock (date and time) are displayed
depending on the User setup programming for
MEMOSKOP. The real time clock is active.
NO: The real time clock is switched off. The User setup
programming will not affect it.
Display of rotated
images
Yes: By pressing the image rotation key, the current
image will be recalculated and displayed.
NO: The rotated image will not be recalculated by press-
ing the image rotation key. For SIREMOBIL Compact
always set to YES.
Monitor zoom
Yes: When pressing the key for MEMOSKOP - Zoom,
the image will be displayed enlarged.
NO: The MEMOSKOP - Zoom key has no function.
For SIREMOBIL Compact set to NO.
Monitor split
Yes: Horizontal or vertical image split possible.
No: No horizontal or vertical image split possible.
Program Yes.
Monitor text
Yes: Superimposes text.
No: No text superimposition.
For SIREMOBIL Compact always set to Yes.
Maximum image numbers
The number of stored images can be programmed.
Program the value to 700 for MEMOSKOP C and to 900
for MEMOSKOP SUB. For MEMOSKOP C-E 100 program
the value to 100. For MEMOSKOP CE, this parameter is
not used.
Product name
Allows you to change the MEMOSKOP name, displayed
during system booting.
Do not change this for SIREMOBIL Compact!
Diagnostics
Allows you to invoke test routines for repair by the manu-
facturer of the MEMOSKOP.
The MEMOSKOP can be reset via "Load factory default,"
as required. Subsequently you have to program accord-
ingly. The same applies if you replace the MEMOSKOP.
Function:
Comments: