Prerequisites
1 - 9
Siemens AG
SPR2-130.061.02
Page 9 of 10
SIREMOBIL Compact
Medical Solutions
Rev. 11
09.04
CS PS 24
System Manual
Back stop plate at the C-arm horizontal support
1
For work that requires removal of the end stop (see 3/Fig. 2), observe the warning above.
Removing the back stop plate
•
Lock the horizontal brake.
•
Remove the rubber cover (adhesive closure) underneath the horizontal support
(see 1/Fig. 2 and 1/Fig. 3).
•
Remove the caps (2/Fig. 2).
•
Remove the three screws (2 / Fig. 2).
•
Remove the back stop plate (3/Fig. 2). Pay attention to the coupling part set on the shaft
of the handle mechanism (2/Fig. 3).
Installing the back stop plate
•
Place the coupling part (2/Fig. 2) back on the shaft.
•
Remount the back stop plate (3/Fig.1) and fasten it with the screws.
•
Reinstall the rubber cover (1/Fig. 1) and close it (adhesive closure).
•
Check the function of the horizontal brake.
Risk of physical injury and damage to property!
When performing service work on the horizontal slide of the basic
unit (for example, when performing repair work on the brake of
the C-arm horizontal travel) it is possible that the C-arm may slide
out of its position and drop onto the floor after the end stop has
been removed.
Before dismantling the end stop, the horizontal travel brake must
be locked and the horizontal slide must be mechanically locked to
prevent it from sliding out. For example, the basic unit can be
positioned in the room in such a way that the C-arm mechanically
touches one of the walls.
Fig. 2
Fig. 3
WARNING
2
1
3
2
1