SIREMOBIL Compact
SPR2-130.061.02
Page 12 of 30
Siemens AG
System Manual
Rev. 11
09.04
CS PS 24
Medical Solutions
5 - 12
Adjustments / Programs
•
Select I.I. survey format.
•
During fluoroscopy, adjust the X-Iris by opening and closing it with the X-Iris keys so that
the collimator blades are still visible along the edge of the I.I.
•
Stop fluoroscopy.
•
Select "2. II. - Full Format Position" on the service PC and click on "Execute".
•
Close the status window by clicking on the OK button.
•
For the query "Do you want to reset the unit," click "No".
•
Switch to zoom format.
•
During fluoroscopy, adjust the X-Iris by opening and closing the X-Iris keys so that the
collimator blades are still visible along the edge of the I.I.
•
Stop fluoroscopy.
•
Select "3. II. - Zoom Format Position" on the service PC and click on "Execute".
•
Close the status window by clicking on the OK button.
•
For the query "Do you want to reset the unit," click "No".
•
Open the X-Iris completely.
•
Select "4. CEX Position" on the service PC and click on "Execute".
•
Close the status window by clicking on the OK button.
•
For the query "Do you want to reset the unit," click "No".
•
Close the X-Iris completely.
•
Release fluoroscopy. The X-Iris must close between 4 - 4.5 cm.
•
Stop fluoroscopy.
•
Select "5. CLOSED POSITION" on the service PC and click on "Execute".
•
Close the status window by clicking on the OK button.
•
For the query "Do you want to reset the unit," click "NO".
•
Select "8. Init Off & Store Positions" on the service PC and click on "Execute".
•
Close the status window by clicking on the OK button.
•
For the query "Do you want to reset the unit," click "Yes".
•
Switch the SIREMOBIL Compact OFF and ON.
•
Check the X-Iris again in survey format and zoom format. The collimator blades must still
be visible on the edge of the image. If necessary, repeat the adjustment.
•
Stop fluoroscopy.
Concluding steps
5
•
Return to the main menu on the service PC with the ESC key.
•
Select the Data, Backup... menu.
•
Select Parameter in the backup window.
•
Click on "Backup". At "Remarks for Parameter" enter "X-Iris set".