SIREMOBIL Compact
SPR2-130.061.02
Page 18 of 30
Siemens AG
System Manual
Rev. 11
09.04
CS PS 24
Medical Solutions
5 - 18
Adjustments / Programs
•
Respond to the query "Do you want to reset the unit" by clicking "No". The slot dia-
phragm remains closed.
Zoom format, slot diaphragm closed
•
The zoom format is selected automatically.
•
Apply fluoro briefly.
•
Open and close the slot diaphragm display via the keys.
•
Align the lines shown on the monitor with the front edges of the slot diaphragm blades.
•
In the window "Available Functions:" select "Slot Diaphragm Close Pos - I.I. Zoom For-
mat" and click "Execute".
•
Close the status window by clicking on the OK button.
•
Respond to the query "Do you want to reset the unit" by clicking "No". The slot dia-
phragm opens automatically (zoom format).
Zoom format, slot diaphragm open
•
Apply fluoro briefly.
•
Open and close the slot diaphragm display via the keys.
•
Align the lines shown on the monitor with the front edges of the slot diaphragm blades.
•
In the window "Available Functions:" select "Slot Diaphragm Open Pos - I.I. Zoom For-
mat" and click "Execute".
•
Close the status window by clicking on the OK button.
•
Respond to the query "Do you want to reset the unit" by clicking "No".
Enable circle mask
•
In the "Available Functions:" window select "Enable Circle Mask" and click on "Execute".
Close the status window by clicking on the OK button.
For the query "Do you want to reset the unit", click on "No".
Save values and end adjustment
•
In the "Available Functions:" window select "INIT OFF & STORE VALUES" and click on
"Execute". Close the status window by clicking on the "OK" button.
For the query "Do you want to reset the Unit", click on "Yes".
Concluding tasks
5
•
Before exiting the service program, save the new parameter values to disk
(Service PC, Data menu, Backup).